MRI Scanner’s Vibration Isolation: Experimental Measurements, Analysis Techniques and Analytical Models
Blekinge Tekniska Högskola, Fakulteten för teknikvetenskaper, Institutionen för tillämpad signalbehandling.
2017 (engelsk)Rapport (Annet vitenskapelig)
Abstract [en]

Modern Magnetic Resonance Imaging (MRI) scanners employ techniques for faster switching of currentsin the gradient coils. The aim is to improve the imaging quality and/or shorter scanning time at thecost of further escalating the associated vibration and noise excited by the Lorentz forces in the gradientcoil. These developments necessitate the employment of effective vibration isolation measures, both priorand post installation, for which a comprehensive analysis of the vibration transfer paths is essential. Such ananalysis is presented in this paper for an operational MRI scanner. The vibration transfer paths are studiedboth analytically and experimentally. Based on the spectral analysis results, improvements in the existingvibration isolation mechanism are discussed.

sted, utgiver, år, opplag, sider
2017. , s. 47
HSV kategori
Identifikatorer
URN: urn:nbn:se:bth-13852OAI: oai:DiVA.org:bth-13852DiVA, id: diva2:1070281
Tilgjengelig fra: 2017-01-31 Laget: 2017-01-31 Sist oppdatert: 2017-02-06bibliografisk kontrollert
Inngår i avhandling
1. Measurements, Analysis Techniques and Experiments in Sound and Vibration: Applied to Operational MRI Scanners and in Remote Laboratories.
Åpne denne publikasjonen i ny fane eller vindu >>Measurements, Analysis Techniques and Experiments in Sound and Vibration: Applied to Operational MRI Scanners and in Remote Laboratories.
2017 (engelsk)Doktoravhandling, med artikler (Annet vitenskapelig)
Abstract [en]

High quality noise and vibration measurements outside of a laboratory environment on real life structures and applications are not trivial. True boundary and operating conditions enforce unique challenges on the measurements. Measurements in hazardous situations such as high magnetic fields, and high temperature environments, etc., where ordinary measurement equipment and methods may not be employed, require further precautions. Post measurements objectives such as analysis, design and strategic decisions, e.g., control, rely heavily on the quality and integrity of the measurements (data).

The quality of the experimental data is highly correlated with the on-field expertise. Practical or hands-on experience with measurements can be imparted to prospective students, researchers and technicians in the form of laboratory experiments involving real equipment and practical applications. However, achieving expertise in the field of sound and vibration measurements in general and their active control in particular is a time consuming and expensive process. Consequently most institutions can only afford a single setup, resulting in the compromise of the quality of expertise.

In this thesis, the challenges in the field of sound and vibration measurements in high magnetic field are addressed. The analysis and measurement of vibration transferred from an operational magnetic resonance imaging (MRI) scanner to adjacent floors is taken as an example. Improvised experimental measurement methods and custom-made frequency analysis techniques are proposed in order to address the challenges and study the vibration transfer. The methods may be extended to other operational industrial machinery and hazardous environments. To encourage and develop expertise in the field of acoustic/vibration measurements and active noise control on practical test beds, remotely controlled laboratory setups are introduced. The developed laboratory setup, which is accessed and controlled via the Internet, is the first of its kind in the active noise control and acoustic measurements area. The laboratory setup can be shared and utilized 24/7 globally, thus reducing the associated costs and eliminating time restrictions.

sted, utgiver, år, opplag, sider
Karlskrona: Blekinge Tekniska Högskola, 2017. s. 251
Serie
Blekinge Institute of Technology Doctoral Dissertation Series, ISSN 1653-2090 ; 3
Emneord
Active Noise Control, Remote Laboratories, Sound and Vibration Measurements, Vibration Analysis, Vibration Transmission
HSV kategori
Identifikatorer
urn:nbn:se:bth-13821 (URN)978-91-7295-336-9 (ISBN)
Disputas
2017-02-22, 10:00 (engelsk)
Opponent
Veileder
Tilgjengelig fra: 2017-02-01 Laget: 2017-01-24 Sist oppdatert: 2017-10-25bibliografisk kontrollert

Open Access i DiVA

fulltext(18762 kB)653 nedlastinger
Filinformasjon
Fil FULLTEXT01.pdfFilstørrelse 18762 kBChecksum SHA-512
59e7b65dbf711017c3dc3683960242334e518c887667995f69671b320bd4c6411a3e3bc85631355f6db9b805a73845e320094df6c5ba5ad3bc4eed1807d1c635
Type fulltextMimetype application/pdf

Personposter BETA

Imran, khan

Søk i DiVA

Av forfatter/redaktør
Imran, khan
Av organisasjonen

Søk utenfor DiVA

GoogleGoogle Scholar
Totalt: 653 nedlastinger
Antall nedlastinger er summen av alle nedlastinger av alle fulltekster. Det kan for eksempel være tidligere versjoner som er ikke lenger tilgjengelige

urn-nbn

Altmetric

urn-nbn
Totalt: 283 treff
RefereraExporteraLink to record
Permanent link

Direct link
Referera
Referensformat
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Annet format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annet språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf